NIM Bin Power Supply Model 8000
Specification
- Input: 200-258 Vac, 47-63 Hz
- Maximum dc output power from 0 to 50 °C : 160 W
- Regulation: <± 0.1% (typically ±0.05%) for ±12 V and ±24 V, and <± 0.2% ( typically <± 0.1% ) for ± 6V
- Output impedance: < 0.3 Ω at any frequency up to 100 KHZ for the dc output
- Temperature cefficient: <0.02%/°C, 0 to 60 °C
- Noise and ripple: < 3 mV peak – to – peak for any output as observed on a 50 MHZ bandwidth oscilloscope
- Voltage Adjustment: ±2% minimum range
NIM Bin Power Supply Model 8010
NIM Bin Power Supply Calibration
Set output voltages of ±6, ± 12 and ± 24 V using internal potentiometer
NIM Bin Power Supply Delivery Test
- Power test
- Stability test
- Getting module inside the rack and check the connection of the module’s connectors with instrument connector
- The equipped room with an air conditioning for constant temperature and humidity
Installation Requirements
- Laboratory
- A table for placing device
- Power cable
- Module for testing the rack
- The equipped room with an air conditioning for constant temperature and humidity
- Power
- Voltmeter or oscilloscope
NIM Bin Power Supply Operating Manual
- Connect the power cable to the socket
- Check the fuses on the back of the device
- Set the front panel switch to ON
- Check the voltages of the power supply connectors with a voltmeter
- Set the front panel switch to ON
- Insert the required modules into the rack and ensure the correct connection of the connectors
- Set the front panel switch to ON
Long Term Warranty And After Sales Service
The warranty service includes the repair or service of the instrument and the free provision of services for replacing parts and repair wages for one year, but the warranty does not apply on the following cases:
- Transport damages, electrical fluctuations, fire or heat, contact or penetration of water and corrosive chemicals, intense dust, lightning, natural disasters.
- Instruments manipulated or repaired by persons other than the company’s representatives.
- Any kind of manipulation or damage on installed holograms on the instrument.
- Any displacement after installation without informing the company.
- The warranty does not cover consumables.
- After sales service is for 10 years with payment.