IAPX100M-Ray Tube

IAPX120 is a medical X-ray generator for producing a beam of high intensity X-rays with small focal spot and high stability which ensures uniform beam intensities and dose rate throughout its Fan-Shaped beam. Stable voltage and electrical power applied to X-ray tube guarantees stable dose exposure at high quality images in digital radiography systems.

 

Technical Specification

  • Tube type: glass tube, tungsten target, beam filter
  • Focal spot:0.5 mm
  • Beam filter: 2 mm thick 6061 AL, 20 mm oil, 1.8 mm glass
  • Beam geometry: Asymmetrical fan up 80 × 30
  • Input power: 200-240 V ac, 50 Hz, 2 A maximum
  • X-ray tube output voltage: Nominal X-ray tube voltage is adjustable between 80 KV to 100 KV with 5 KV step
  • X-ray tube current: 0.2 mA to 1.2 mA, adjustable
  • X-ray tube power: 100 W, continuous mode (can be increased)
  • Voltage regulation:

1. Line: ±1% for a ±10% input line change of nominal input line voltage

2. Load: ± 0.1 for a 0.2 mA to 1.2 mA load change

3. Voltage accuracy: Voltage measured across the X-ray tube is within ± 2% of the programmed value

  • Current regulation:

1. Line: ±1% for a ±10% input line change of nominal input line voltage

2. Load: 100-160 for 0.5% KV, 0.1 mA to 1.2 mA

3. Current accuracy: Current measured through the X-ray tube is within ± 2% of the programmed value

4. Current rise time: < 200 ms from 10% to 90% of rated output

  • Digital interface: Rs232-/USB digital interface selectable ports
  • Control software: A demo GUI for engineering evaluations will be provided for the RS23-/USB digital interface and encoded command port for customized software.
  • Software: Graphical version
  • Operating temperature: 0 to +40 C°
  • Storage temperature: -40 to +70 C°
  • Humidity: 10% to 95% relative humidity, non-condensing
  • Dimensions: 600 mm × 390 mm × 160 mm
  • Weight: 40 Kg
  • Installation orientation: Can be mounted in any orientation
  • Beam orientation: Beam divergence angle: 80 degrees while 20 degrees rotated centrally toward central part of the tube body
  • X-ray leakage: Not to be greater than 1 mm R/hr at 30 cm outside the external surface
  • Accessories: CD software, serial cable and RS232, user manual, external collimator
  • Accessories

1. USB connection cable

2. RS232-connection cable

3. User manual

4. Software

 

Applications

  • X-ray imaging
  • Nondestructive testing
  • Food inspection
  • Security inspection
  • Bone mineral densitometry

 

Long Term Warranty And After Sales Service

The warranty service includes the repair or service of the instrument and the free provision of services for replacing parts and repair wages for one year, but the warranty does not apply on the following cases:

  • Transport damages, electrical fluctuations, fire or heat, contact or penetration of water and corrosive chemicals, intense dust, lightning, natural disasters.
  • Instruments manipulated or repaired by persons other than the company’s representatives.
  • Any kind of manipulation or damage on installed holograms on the instrument.
  • Any displacement after installation without informing the company.
  • The warranty does not cover consumables.
  • After sales service is for 10 years with payment.